Braun IRT6020 Spezifikationen Seite 7

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 59
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 6
10 11
Errors and troubleshooting
Product speci cations
Displayed temperature range: 34 – 42.2 °C (93.2 – 108 °F)
Operating ambient temperature range: 10 – 40 °C (50 – 104 °F)
Storage temperature range: -25 – 55 °C (-13 – 131 °F)
Operating and storage relative humidity: 10-95% RH (non condensing)
Display resolution: 0.1 °C or °F
Accuracy for displayed temperature range Maximum Laboratory Error
35 – 42 °C (95 – 107.6 °F): ±0.2 °C (±0.4 °F)
Outside this range: ±0.3 °C (±0.5 °F)
Clinical repeatability: ±0.14 °C (±0.26 °F)
Battery life: 2 years / 600 measurements
Service life: 5 years
This thermometer is speci ed to operate at 1 atmospheric pressure or at altitudes with an atmospheric
pressure up to 1 atmospheric pressure (700 - 1060 hPa).
See Instruction for use
Equipment with
type BF applied parts
10 °C
40 °C
Operating temperature
Keep it dry
Storage temperature
Subject to change without notice.
This appliance conforms to the following standards:
Standard Reference Edition Title:
EN 12470-5: 2003 Clinical thermometers – Part 5: Performance of infra-red ear thermometers (with
maximum device).
EN 60601-1: 2006 Medical electrical equipment – Part 1: General requirements for basic safety and essential
performance.
EN ISO 14971: 2012 Medical devices – Application of risk management to medical devices.
EN ISO 10993-1: 2009 Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk
management process.
EN 60601-1-2: 2007 Medical electrical equipment – Part 1-2: General requirements for basic safety and
essential performance – Collateral standard: Electromagnetic compatibility – Requirements and tests.
EN 980: 2008 Symbols for use in labeling of medical devices.
EN 1041: 2008 Information supplied by the manufacturer of medical devices.
EN 60601-1-11: 2010 Medical electrical equipment – Part 1-11: General requirements for basic safety and
essential performance – Collateral standard: Requirements for medical electrical equipment and medical
electrical systems used in the home healthcare environment.
This product conforms to the provisions of the EC directive 93/42/EEC.
MEDICAL ELECTRICAL EQUIPMENT needs special precautions regarding EMC. For detailed description of
EMC requirements please contact your authorized local Service Centre (See guarantee card).
Portable and mobile RF communications equipment can a ect MEDICAL ELECTRICAL EQUIPMENT.
Please do not dispose of the product in the household waste at the end of it useful life.
To protect the environment, dispose of empty batteries at appropriate collection sites according to
national or local regulations.
Guarantee
Consumer Card available on our website at www.hot-europe.com/after-sales
Please see last page of this manual to  nd the contact for the Kaz Authorized Service Center in your country.
UK only: This guarantee in no way a ects your rights under statutory law.
English
Error Message Situation Solution
No lens  lter is attached. Attach new, clean lens  lter.
The thermometer probe was not
positioned securely in the ear.
An accurate measurement
was not possible.
POS = position error
Take care that the positioning of the
probe is correct and remains stable.
Replace lens  lter and reposition.
Press Start button to begin a new
measurement.
Ambient temperature is not within
the allowed operating range
(10 – 40 °C or 50 – 104 °F).
Allow the thermometer to remain for
30 minutes in a room where
the temperature is between 10
and 40 °C or 50 and 104 °F .
Temperature taken is not within
typical human temperature range
(34 – 42.2 °C or 93.2 – 108 °F).
HI = too high
LO = too low
Make sure the probe tip and lens are
clean and a new, clean lens  lter is
attached. Make sure the thermometer
is properly inserted. Then, take a new
temperature.
System error – self-check display
ashes continuously and will not be
followed by the ready beep and the
ready symbol.
If error persists,
If error still persists,
Wait 1 minute until the thermometer
turns o automatically, then turn
on again.
... reset the thermometer by removing
the batteries and putting them back in.
... please contact the Service Centre.
Battery is low, but thermometer will
still operate correctly.
Insert new batteries.
Battery is too low to take correct
temperature measurement.
Insert new batteries.
Seitenansicht 6
1 2 3 4 5 6 7 8 9 10 11 12 ... 58 59

Kommentare zu diesen Handbüchern

Keine Kommentare