Guidance and manufacturer’s declaration – electromagnetic emissions
The ME equipment is intended for use in the electromagnetic environment specied below.
The customer or the user of the ME equipment should ensure that it is used in such an environment.
Emissions Test Compliance Electromagnetic environment – guidance
RF Emissions
CISPR 11
Group 1
The ME equipment uses RF energy only for its internal function. Therefore, its RF emissions are
very low and are not likely to cause any interference in nearby electronic equipment.
RF Emissions
CISPR 11
Class B Complies
Harmonic emissions IEC
61000-3-2
Not Applicable
The ME equipment is
solely battery powered.
Voltage uctuations /
icker emissions
Not Applicable
Non-Life Support Equipment Separation Distance Calculation (3Vrms / 3V/m compliance)
Rated maximum output power of
transmitter (W)
Separation distance according to frequency of transmitter (m)
150 kHz to 80 MHz in ISM bands 80 MHz to 800 MHz 800 MHz to 2.5 GHz
0.01 0.12 0.12 0.23
0.1 0.37 0.37 0.74
1 1.17 1.17 2.33
10 3.69 3.69 7.38
100 11.67 11.67 23.33
Guidance and manufacturer’s declaration – electromagnetic immunity
The ME equipment is intended for use in the electromagnetic environment specied below.
The customer or the user of the ME equipment should ensure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment – guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
±6kV Contact
±8kV Air
Complies
Floors should be wood, concrete, or ceramic tile. If
oors are covered with synthetic material, the
relative humidity should be at least 30%.
Radiated RF
IEC 61000-4-3
3 V/m 80MHz to 2.5GHz Complies
Field strengths outside the shielded location from
xed RF transmitters, as determined by an
electromagnetic site survey, should be less than 3
V/m.
Interference may occur in the vicinity of equipment
marked with the following symbol:
Separation distance calculation provided above. If a
known transmitter is present the specic distance
can be calculated using the equations.
Conducted RF
IEC 61000-4-6
3Vrms 150kHz to 80MHz
Not Applicable
(no electrical cabling)
Electrical fast transient
IEC 61000-4-4
±2kV power line
±1kV I/O lines
Not Applicable
The ME equipment
is solely battery powered.
Surge
IEC 61000-4-5
±1kV dierential
±2kV common
Not Applicable
Power frequency magnetic eld
IEC 61000-4-8
3 A/m Complies
Power frequency magnetic elds should be at levels
characteristic of a typical location in a typical
commercial or hospital environment.
Voltage dips, short interruptions
and voltage variations on power
supply input lines
IEC 61000-4-11
>95% dip 0.5 cycle
60% dip 5 cycles
70% dip 25 cycles
95% dip 5 sec.
Not Applicable
The ME equipment
is solely battery
powered.
d = [
3,5
] P
V
1
d = [
3,5
] P
E
1
d = [
7
] P
E
1