Special technical features
The configuration of the electrode connection provides protection
against a wrong polarity connection of the electrode cable.
This new type of plug connection system ensures correct polarity
of the electrodes at all times.
A high stimulation voltage is associated with an unusually wide
range of adjustment with extremely small stimulation electrodes.
The Stimuplex
®
HNS 11 generates a negative, current-stabilized
square pulse. In contrast to conventional instruments of this type,
the pulse is shaped at both slopes by extremely fast active pulse
drivers. An output amplifier specially designed for this application
has an extraordinarily wide dynamic range and produces repro-
ducible settings even below 0.1 mA.
Use of the latest microcomputer technology has made it possible
to integrate performance characteristics which until now could
only be provided in large systems. For example: selection of cur-
rent display mode, set and actual value display, display of battery
charge, error messages.
An internal quartz time base ensures precise pulse widths and
frequencies.
12
Technical data
Due to its high maximum output voltage the nerve stimulator cov-
ers a wide range of skin and needle impedance. This is especially
important when using stimulation needles with an extremely small
conductive area (pin point electrode tip) such as Stimuplex
®
D and
Contiplex
®
D needles.
Type: Stimuplex
®
HNS 11 nerve stimulator
Instrument type ................ BF
Battery .................................9 V
Power consumption ......... 3.3 mA
Stimulation current .........max. 5 mApp / 0 - 12 k
Stimulation voltage ........max. 65 Vpp
Stimulation frequency ....1 Hz / 2 Hz
Measuring tolerance .......Adjustment control display =3% (set point)
Flowing current display = 2% (actual value)
based on set mA max values (5 mA or 1 mA)
Weight .................................250 g